Medlop IP1 Infusion Pump

28,000.00

Main Features

☆ HD LCD Display, high capacity words, friendly user interface, dynamically display working status.

☆ Audible and visual alarm for occlusion, empty, low battery, end of infusion, door open, wrong setting, etc., which gain patents.

☆ Compatible with any brands of infusion sets after correct calibration.

☆ Preset solution volume to greatly reduce the workload of nurses.

☆ Work mode: ml/h and drop/min can switch freely.

☆ Three levels of occlusion: high, medium and low.

☆ Purge function.

☆ KVO (keep-vein-open) automatically opens as infusion is completed, KVO rate is 1-5ml/h(1ml/h step).

☆ Power Source: AC100—240V, 50/60Hz; Internal Battery.

☆ Automatically record the settings of last infusion.

Specifications

Infusion Rate ml/h mode:1~1200ml/h(1ml/h step)

drop/min mode:1~266drop/min(1drop/min step)

Flow Rate Accuracy ml/h mode:Within±5%(after correct calibration)

drop/min mode:Within±3%(after correct calibration)

Purge Rate 100mh/h—1200ml/h(can be calibrated)
Mechanical Precision Within±2%
Maximum Infusion Pressure 150kPa
Alarm Pressure High 800mmHg±200mmHg(106.7kPa±26.7kPa)

Medium 500mmHg±100mmHg(66.7kPa±13.3kPa)

Low 300mmHg±100mmHg(40.7kPa±13.3kPa)

Audible and Visible Alarms There are bubbles, Tube is blocked, Close the door, IV set Mismatched, Battery running out, Infusion finish, Equipment failure, Abnormally Operation, Software Exception, Low battery, Forget the operations etc
Transport & Storage Condition Temperature -20℃~+55℃

Relative Humidity ≤95%

Atmospheric Pressure 50.10KPa—106.0KPa

Infusion Set Compatible with any brand of infusion sets after correct calibration
Infusion maximum value (under single fault condition) 1ml
Forbidden Not support blood Infusion
Dimension 140mm(L)×157mm(W)×220mm(H)
Net Weight 1.8kg
Package Reference 1pcs  Carton:2.8KG,

33cm(L)×28cm(W)×26cm(H)