Main Features
☆ HD LCD Display, high capacity words, friendly user interface, dynamically display working status.
☆ Audible and visual alarm for occlusion, empty, low battery, end of infusion, door open, wrong setting, etc., which gain patents.
☆ Compatible with any brands of infusion sets after correct calibration.
☆ Preset solution volume to greatly reduce the workload of nurses.
☆ Work mode: ml/h and drop/min can switch freely.
☆ Three levels of occlusion: high, medium and low.
☆ Purge function.
☆ KVO (keep-vein-open) automatically opens as infusion is completed, KVO rate is 1-5ml/h(1ml/h step).
☆ Power Source: AC100—240V, 50/60Hz; Internal Battery.
☆ Automatically record the settings of last infusion.
Specifications
Infusion Rate | ml/h mode:1~1200ml/h(1ml/h step)
drop/min mode:1~266drop/min(1drop/min step) |
Flow Rate Accuracy | ml/h mode:Within±5%(after correct calibration)
drop/min mode:Within±3%(after correct calibration) |
Purge Rate | 100mh/h—1200ml/h(can be calibrated) |
Mechanical Precision | Within±2% |
Maximum Infusion Pressure | 150kPa |
Alarm Pressure | High 800mmHg±200mmHg(106.7kPa±26.7kPa)
Medium 500mmHg±100mmHg(66.7kPa±13.3kPa) Low 300mmHg±100mmHg(40.7kPa±13.3kPa) |
Audible and Visible Alarms | There are bubbles, Tube is blocked, Close the door, IV set Mismatched, Battery running out, Infusion finish, Equipment failure, Abnormally Operation, Software Exception, Low battery, Forget the operations etc |
Transport & Storage Condition | Temperature -20℃~+55℃
Relative Humidity ≤95% Atmospheric Pressure 50.10KPa—106.0KPa |
Infusion Set | Compatible with any brand of infusion sets after correct calibration |
Infusion maximum value (under single fault condition) | 1ml |
Forbidden | Not support blood Infusion |
Dimension | 140mm(L)×157mm(W)×220mm(H) |
Net Weight | 1.8kg |
Package Reference | 1pcs Carton:2.8KG,
33cm(L)×28cm(W)×26cm(H) |