Medlop SP1 Syringe Pump

24,000.00

Main Features

☆ HD LCD Display, high capacity words, friendly user interface, dynamically display working status.

☆ Audible and visual alarm for occlusion, empty, near empty, low battery, end of infusion, syringe loose, wrong setting, etc.

☆ Compatible with 5ml, 10ml, 20ml, 30ml, 50/60ml syringe of any brands.

☆ Preset solution volume to greatly reduce the workload of nurses.

☆ Three working modes: Rate mode, Time Volume mode, Dosage Weight mode.

☆ Three levels of occlusion: high, medium and low.

☆ Purge and Bolus function.

☆ KVO (keep-vein-open) automatically opens as injection is completed, KVO rate is 0.1-5ml/h(0.1ml/h step).

☆ Freely Stackable: freely stack one syringe pump onto another to provide multiple solutions, which have a wide range of clinical applications.

☆ Power Source: AC100—240V, 50/60Hz; Internal Battery, DC12V car charge.

☆ One-key operation makes setup easy and simple.

☆ Syringe plunger grabble detector, operable with one hand in a No-Germ Environment.

☆ Automatically record the settings of last injection.

Specifications

Infusion Rate 50/60ml Syringe: 0.1ml/h~1500ml/h(0.1ml/h step)

30ml Syringe: 0.1ml/h—900ml/h(0.1ml/h step)

20ml Syringe:0.1ml/h—600ml/h(0.1ml/h step)

10ml Syringe 0.1ml/h—300ml/h(0.1ml/h step)

Rate Accuracy Within±3%(after correct calibration)
Maximum Stackable 8 pcs
Purge Rate 50/60ml Syringe:1500ml/h

30ml Syringe:900ml/h

20ml Syringe:600ml/h

10ml Syringe:300ml/h

Mechanical Accuracy Within±2%
Bolus Rate 50/60ml Syringe:1200ml/h

30ml Syringe:720ml/h

20ml Syringe:480ml/h

10ml Syringe:240ml/h

Volume Limit 0.1ml—9999.9ml(0.1ml step)
Total Injection 0.1ml—9999.9ml(0.1ml step)
Maximum Infusion Pressure 150kPa
Occlusion High 800mmHg±200mmHg(106.7kPa±26.7kPa)

Medium 500mmHg±100mmHg(66.7kPa±13.3kPa)

Low 300mmHg±100mmHg(40.7kPa±13.3kPa)

Dimension 280mm(L) x 210mm(W) x 130mm(H)
Net Weight 2.2kg
Package Reference 1pc Carton:3.2KG,36cm(L) x 29cm(W) x 18cm(H)
Syringe Compatible with 10ml,20ml,30ml,50/60ml syringe of any brands after correct calibration